CBET Exam Prep Free practice test →

Free CBET Practice Questions

10 free, exam-style Certified Biomedical Equipment Technician (CBET) practice questions with answers and explanations. No signup required. Work through them below, then take the full free CBET practice test to study every exam domain.

Question 1

While verifying the performance of an invasive arterial blood pressure monitor, a technician observes the dicrotic notch on the displayed pressure waveform. This notch corresponds to:

  1. Opening of the aortic valve
  2. Closure of the aortic valve
  3. Closure of the mitral valve
  4. The peak of ventricular ejection
Show answer & explanation

Correct answer: B - Closure of the aortic valve

Question 2

A technician is performing an electrical safety test on a device that makes a direct electrical connection to the heart, which requires a Type CF applied part. Under IEC 60601-1, the maximum allowable patient leakage current for a Type CF applied part in the normal (no-fault) condition is:

  1. 10 µA
  2. 50 µA
  3. 100 µA
  4. 500 µA
Show answer & explanation

Correct answer: A - 10 µA

Question 3

A component in a DC circuit has 12 volts measured across it while 3 amperes of current flow through it. Applying Ohm's law, the component's resistance is:

  1. 0.25 Ω
  2. 15 Ω
  3. 4 Ω
  4. 36 Ω
Show answer & explanation

Correct answer: C - 4 Ω

Question 4

The hypoxic guard (oxygen ratio controller) on an anesthesia machine is designed to prevent the clinician from delivering a fresh gas mixture that contains:

  1. Too high a concentration of oxygen
  2. Too low a concentration of oxygen
  3. Insufficient total fresh gas flow
  4. An excessive concentration of anesthetic vapor
Show answer & explanation

Correct answer: B - Too low a concentration of oxygen

Question 5

A pulse oximeter determines oxygen saturation by comparing how oxygenated and deoxygenated hemoglobin absorb two specific wavelengths of light. The wavelengths used are:

  1. Two separate infrared wavelengths
  2. Ultraviolet and blue light
  3. Green and yellow light
  4. Red and infrared light
Show answer & explanation

Correct answer: D - Red and infrared light

Question 6

On a monopolar electrosurgical unit, the return electrode monitoring (REM) system continuously checks the quality of contact between the dispersive (return) pad and the patient. Its PRIMARY purpose is to:

  1. Prevent burns at the return electrode site
  2. Increase the active electrode's cutting power
  3. Reduce the smoke produced during coagulation
  4. Automatically alternate between cut and coagulation
Show answer & explanation

Correct answer: A - Prevent burns at the return electrode site

Question 7

A technician investigates a bedside monitor showing a steady, fine, regular 60 Hz interference on every ECG lead; the same pattern is present on several other monitors in the same unit. The MOST likely cause is:

  1. A single dried-out ECG electrode
  2. Patient shivering or muscle tremor
  3. Interference from nearby AC power lines
  4. An intermittently broken patient lead wire
Show answer & explanation

Correct answer: C - Interference from nearby AC power lines

Question 8

A defibrillator set to 200 J is discharged several times into a defibrillator/energy analyzer and delivers only about 150 J each time. Given that delivered energy should fall within roughly ±15% of the set value, the technician should conclude that:

  1. The reading falls within the normal tolerance allowed for defibrillators
  2. The analyzer should be set to monophasic mode
  3. Some energy loss is expected and no action is needed
  4. The device is faulty and must be removed from service
Show answer & explanation

Correct answer: D - The device is faulty and must be removed from service

Question 9

A volumetric infusion pump is returned to the clinical engineering shop three times for the same reported 'malfunction.' Each time it passes all functional and electrical-safety tests, and review shows that staff had repeatedly loaded the IV administration set incorrectly. This situation is BEST classified as:

  1. A use error
  2. An intermittent hardware fault
  3. A latent device failure
  4. A manufacturing defect requiring a recall
Show answer & explanation

Correct answer: A - A use error

Question 10

To reduce cybersecurity risk, a hospital places its networked infusion pumps and patient monitors on a dedicated VLAN that is isolated from the general staff and guest networks. This practice is BEST described as:

  1. Data encryption
  2. Network segmentation
  3. Load balancing
  4. Multifactor authentication
Show answer & explanation

Correct answer: B - Network segmentation

Ready for the real thing?

Practice hundreds more CBET questions with instant scoring, weak-area drills, and full exam simulations.

Start the free practice test See pricing